Histotripsy-induced bactericidal activity correlates to size of cavitation cloud in vitro
  • +16
  • P A Ambekar,
  • Y N Wang,
  • T Khokhlova,
  • G Thomas,
  • P Rosnitskiy,
  • K Contreras,
  • D F Leotta,
  • A D Maxwell,
  • M Bruce,
  • S Pierson,
  • S Totten,
  • Y N Kumar,
  • J Thiel,
  • K Chan,
  • W C Liles,
  • E P Dellinger,
  • A Adedipe,
  • W Monsky,
  • Thomas Matula
P A Ambekar
Y N Wang
T Khokhlova
G Thomas
P Rosnitskiy
K Contreras
D F Leotta
A D Maxwell
M Bruce
S Pierson
S Totten
Y N Kumar
J Thiel
K Chan
W C Liles
E P Dellinger
A Adedipe
W Monsky
Thomas Matula

Corresponding Author:[email protected]

Author Profile

Abstract

Large abscesses are walled-off collections of pus and bacteria that often do not respond to antibiotic therapy. Standard of care involves percutaneous placement of indwelling catheter(s) for drainage, a long and uncomfortable process with high rehospitalization rates. The long-term goal of this work is to develop therapeutic ultrasound approaches to eradicate bacteria within abscesses as a noninvasive therapeutic alternative. Inertial cavitation induced by short pulses of focused ultrasound (histotripsy) is known to generate lethal mechanical damage in bacteria. Prior studies with Escherichia coli (E. coli) in suspension demonstrated that bactericidal effects increase with increasing peak negative pressure amplitude, treatment time and duty cycle. The current study investigated correlates of bactericidal activity with histotripsy bubble cloud size produced for these treatment parameters at several ultrasound frequencies. Histotripsy was applied to E. coli suspensions in 10-mL sample vials at 810 kHz, 1.2 MHz, or 3.25 MHz for 40 minutes. The resulting cavitation activity in the sample vials was separately observed with high-speed photography. The bubble cloud area was quantified from those images. A linear relationship was observed between log kill and cavitation cloud size (R2 = 0.96), regardless of the acoustic parameters (specifically frequency, pulse duration and pressure amplitude) used to produce the cloud. These findings suggest a role for in-situ treatment monitoring using B-mode imaging.
12 Apr 2024Submitted to TechRxiv
18 Apr 2024Published in TechRxiv