Technical design considerations of a human-scale Talbot-Lau interferometer for dark-field CT
  • +6
  • Manuel Viermetz ,
  • Nikolai Gustschin ,
  • Clemens Schmid ,
  • Jakob Haeusele ,
  • Peter Noël ,
  • Roland Proksa ,
  • Stefan Loescher ,
  • Thomas Koehler ,
  • Franz Pfeiffer
Manuel Viermetz
Author Profile
Nikolai Gustschin
Author Profile
Clemens Schmid
Author Profile
Jakob Haeusele
Author Profile
Peter Noël
Author Profile
Roland Proksa
Author Profile
Stefan Loescher
Author Profile
Thomas Koehler
Author Profile
Franz Pfeiffer
Author Profile

Abstract

Computed tomography (CT) as an important clinical diagnostics method can profit from extension with dark-field imaging, as it is currently restricted to X-rays’ attenuation contrast only. Dark-field imaging allows access to more tissue properties, such as micro-structural texture or porosity. The up-scaling process to clinical scale is complex because several design constraints must be considered. The two most important ones are that the finest grating is limited by current manufacturing technology to a 4.8µm period and that the interferometer should fit into the CT gantry with minimal modifications only. In this work we discuss why an inverse interferometer and a triangular G1 profile are advantageous and make a compact and sensitive interferometer implementation feasible. Our evaluation of the triangular grating profile reveals a deviation in the interference pattern compared to standard grating profiles, which must be considered in the subsequent data processing. An analysis of the grating orientation demonstrates that currently only a vertical layout can be combined with cylindrical bending of the gratings. We also provide an in-depth discussion, including a new simulation approach, of the impact of the extended X-ray source spot which can lead to large performance loss and present supporting experimental results. This analysis reveals a vastly increased sensitivity to geometry and grating period deviations, which must be considered early in the system design process.
Jan 2023Published in IEEE Transactions on Medical Imaging volume 42 issue 1 on pages 220-232. 10.1109/TMI.2022.3207579